Passive neutron area monitor with CR39
|
|
|
|
|
|
|
|
نویسنده
|
Vega-Carrillo H.R. ,Barquero R. ,Mercado G.A.
|
|
منبع
|
international journal of radiation research - 2013 - دوره : 11 - شماره : 3 - صفحه:149 -153
|
|
چکیده
|
Background: in high-intensity, mixed and pulsed neutron fields the use of spectrometers or area monitors with active detectors is useless; in these conditions neutron measuring devices must have a passive detector. here a passive neutron area monitor with cr39 track detector was designed and the response was calculated. materials and methods: the response of a passive neutron area monitor with cr39 track detector has been calculated with the mcnpx code. to increase the detection efficiency a 10b converter was included. the response was calculated using 47 monoenergetic neutron sources. results: a passive neutron area monitor using a cr39 with 10b converter was designed where fluence and h*(10) responses were calculated. conclusion: the shape of the responses is similar to responses reported for neutron monitors with active detectors.
|
|
کلیدواژه
|
Neutron dosimetry ,passive detector ,CR-39 ,Monte Carlo ,response.
|
|
آدرس
|
Universidad Autonoma de Zacatecas, Unidad Académica de Estudios Nucleares, Mexico, Hospital Universitario Rio Hortega, Departamento de Radiología, Spain, Universidad Autónoma de Zacatecas, Unidad Academica de Matematicas, Mexico
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|