   Assessment of patient dose in routine digitalradiography in Iran  
   
نویسنده mohsenzadeh b. ,deevband m.r. ,paydar r. ,ghorbani m.
منبع international journal of radiation research - 2020 - دوره : 18 - شماره : 3 - صفحه:449 -460
چکیده    Background: the present study aimed to investigate patient dose in commonx-ray examinations to estimate effective dose in the digital radiography iniran. materials and methods: entrance surface dose (esd) was measuredbased on applied exposure parameters for the common actual examination;and then effective dose (ed) was calculated by use of pcxmc software. thestudy was conducted on 15358 patients in 85 x-ray rooms; and the necessarydata was collected for five age groups, 0-1 year, 1-5 years, 5-10 years, 10-15years old and adults in each projection. results: the ranges of esd and ed indifferent examinations for all the age groups are 0.02-10.20 mgy and 2.42-378.96 μsv respectively. conclusion: the effective dose as criteria can be usedto reduce patients' doses. the special considerations such as: adequatetraining of imaging staff; updating clinical audits; patient dose considerations;implementation of systematic and regular quality assurance and qualitycontrol programs in medical imaging departments should be taken intoaccount to optimize radiological practices.
کلیدواژه Patient dose; dose reference level; dosimetry in diagnostic radiology;entrance surface dose; effective dose.
آدرس shahid beheshti university of medical sciences, faculty of medicine, biomedical engineering and medical physics department, Iran, shahid beheshti university of medical sciences, faculty of medicine, biomedical engineering and medical physics department, Iran, iran university of medical sciences and health services, department of radiology, Iran, shahid beheshti university of medical sciences, faculty of medicine, biomedical engineering and medical physics department, Iran
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved