   Assessment of basic physical and dosimetric parameters of synthetic single-crystal diamond detector and its use in Leksell Gamma Knife and CyberKnife small radiosurgical fields  
   
نویسنده veselsky t. ,novotny jr j. ,pastykova v.
منبع international journal of radiation research - 2018 - دوره : 16 - شماره : 1 - صفحه:7 -16
چکیده    Background: to determine the basic physical and dosimetric properes of a newsynthec single-crystal diamond detector and its applicaon for relave small fielddosimetry. materials and methods: the pre-irradiaon dose required to stabilizedetector response, dose rate dependence, photon and electron energydependence, temperature dependence and angular dependence of microdiamonddetector response were evaluated. output factors on leksell gamma knifeperfexion and on cyberknife were measured to assess detector feasibility in smallradiaon field dosimetry. for all measurements, the detector was connected tounidos electrometer set to 0 voltage. results: relave output factors measuredon leksell gamma knife perfexion for 4 mm and 8 mm collimators were inagreement with monte carlo reference values from the manufacturer, withdeviaons of 0.3% and 2.1%, respecvely. for cyberknife and fixed circularcollimators, the difference in output factor values did not exceed 2% fromvendor-supplied values, even for the smallest radiaon field with a diameterof 5 mm. conclusion: our results indicate that the microdiamond detector is apromising tool for relave small field dosimetry. for output factor measurementson leksell gamma knife perfexion and cyberknife, the detector can be used withminimal response correcons applied (correcon factors not larger than 2%).
کلیدواژه Small field dosimetry ,synthetic diamond detector ,output factor ,radiosurgery
آدرس czech technical university, faculty of nuclear sciences and physical engineering, department of dosimetry and application of ionizing radiation, Czechia. motol university hospital, oncology clinic, czechia. na homolce hospital, department of medical physics, Czechia, czech technical university, faculty of nuclear sciences and physical engineering, department of dosimetry and application of ionizing radiation, czechia. na homolce hospital, department of medical physics, Czechia. motol university hospital, oncology clinic, Czechia, czech technical university, faculty of nuclear sciences and physical engineering, department of dosimetry and application of ionizing radiation, Czechia. na homolce hospital, department of medical physics, Czechia
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved