Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals
|
|
|
|
|
|
|
|
نویسنده
|
Aghahadi B. ,Zhang Z. ,Zareh S. ,Sarkar S. ,Tayebi P.S.
|
|
منبع
|
international journal of radiation research - 2006 - دوره : 3 - شماره : 4 - صفحه:177 -182
|
|
چکیده
|
Background: the x-ray machines used for radiodiagnosis should meet certain quality assurance (qa) programmes. these are necessary to have good quality radiographs at reasonably low exposure to patients. materials and methods: dose reduction methods in abdomen x-ray examination were carried out in 10 hospitals in tehran. this paper presents the work, which was implemented on 200 patients and evaluated using the entrance skin dose (esd) in the anterior-posterior (ap) abdomen projection measured directly at the center of the x-ray field. in addition, the machine room, and dark room parameters, as well as work practices and repeat rates were studied.results: the quality control (qc) parameters and the esd were evaluated utilizing an anthropologic phantom to define the optimal exposure condition at all hospitals before and after qc. results show that after using the qc parameters and optimization of the exposure conditions, the mean of mas and esd can be decreased by 62% and 65% respectively. conclusion: the quality of the radiographs generally increased. the reported method is easily implemented in any clinical situation where optimization of abdomen radiography is necessary.
|
|
کلیدواژه
|
Quality assurance ,quality control ,X-ray radiography ,diagnostic radiology ,entrance skin dose ,dose reduction.
|
|
آدرس
|
Harbin Institute of Technology, Department of Physics, China. Atomic Energy Organization of Iran, Department of National Radiation Protection, ایران, Harbin Institute of Technology, Department of Physics, China, tehran university of medical sciences tums, Department of Medical Physics, ایران, tehran university of medical sciences tums, Department of Medical Physics, ایران, Heilongjiang University, Department of Physics, China
|
|
پست الکترونیکی
|
b.aghahadi@hit.edu.cn
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|