|
|
evaluation of ultra-low-dose chest computed tomography images in detecting lung lesions related to covid-19: a prospective study
|
|
|
|
|
نویسنده
|
zarei fariba ,jalli reza ,chatterjee sabyasachi ,ravanfar haghighi rezvan ,iranpour pooya ,chatterjee vani vardhan ,emadi sedigheh
|
منبع
|
iranian journal of medical sciences - 2022 - دوره : 47 - شماره : 4 - صفحه:338 -349
|
چکیده
|
Background: the present study aimed to evaluate the effectiveness of ultra-low-dose (uld) chest computed tomography (ct) in comparison with the routine dose (rd) ct images in detecting lung lesions related to covid-19. methods: a prospective study was conducted during april- september 2020 at shahid faghihi hospital affiliated with shiraz university of medical sciences, shiraz, iran. in total, 273 volunteers with suspected covid-19 participated in the study and successively underwent rd-ct and uld-ct chest scans. two expert radiologists qualitatively evaluated the images. dose assessment was performed by determining volume ct dose index, dose length product, and size-specific dose estimate. data analysis was performed using a ranking test and kappa coefficient (κ). p<0.05 was considered statistically significant. results: lung lesions could be detected with both rd-ct and uld-ct images in patients with suspected or confirmed covid- 19 (κ=1.0, p=0.016). the estimated effective dose for the rd-ct protocol was 22-fold higher than in the uld-ct protocol. in the case of the uld-ct protocol, sensitivity, specificity, accuracy, and positive predictive value for the detection of consolidation were 60%, 83%, 80%, and 20%, respectively. comparably, in the case of rd-ct, these percentages for the detection of groundglass opacity (ggo) were 62%, 66%, 66%, and 18%, respectively. assuming the result of real-time polymerase chain reaction as true-positive, analysis of the receiver-operating characteristic curve for ggo detected using the uld-ct protocol showed a maximum area under the curve of 0.78. conclusion: uld-ct, with 94% dose reduction, can be an alternative to rd-ct to detect lung lesions for covid-19 diagnosis and follow-up. an earlier preliminary report of a similar work with a lower sample size was submitted to the arxive as a preprint. the preprint is cited as: https://arxiv.org/abs/2005.03347
|
کلیدواژه
|
covid-19 ,radiation protection ,computed tomography
|
آدرس
|
shiraz university of medical sciences, medical imaging research center, iran, shiraz university of medical sciences, medical imaging research center, iran, ongil, india, shiraz university of medical sciences, medical imaging research center, iran, shiraz university of medical sciences, medical imaging research center, iran, indian institute of science, department of instrumentation and applied physics, india, shiraz university of medical sciences, medical imaging research center, iran
|
پست الکترونیکی
|
emadi.majesti@gmail.com
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|