Uncertainty Analysis of H_p(10)_meas/H_p(10)_del Ratio for TLD-100H at Energy 24-1250 keV
|
|
|
|
|
نویسنده
|
Priharti W. ,Samat S. B. ,Kadir A. B. A.
|
منبع
|
jurnal teknologi - 2013 - دوره : 62 - شماره : 3 - صفحه:115 -118
|
چکیده
|
The h_p(10) of an ideal tld is independent of photon energy. this is to say that at any photon energy, the ratio r of the measured dose h_p(10)_meas to the delivered dose h_p(10)_del is always 1. in practice however the h_p(10) is dependent on the energy and r is not equal to 1. for this reason, icrp has set the lower limit ll and upper limit u_l for r as 0.55≤r≤1.63 for detection limit of 0.1 msv and h_p(10)_del=1 msv. as r is the quantity arises from the measurement processes, the existence of uncertainty of r, i.e. δr is inevitable. in the boundary cases, such as when r is slightly lower than l_l or slightly larger than u_l, δr would serve a useful quantity in decision making either to accept or to reject the value of r. the purpose of the present work is to estimate δr for the tld- 100h for photon energy of 24-1250 kev. the estimation of δr is based on the error propagation method. for the eleven photon energies, this work obtained (a) r in the range of 0.77 to 1.16, (b) δr in the range of ± 0.02 to ± 0.04. the values of r were satisfactory as they complied the icrp limit. the determined δr is considered very small as it is in the order of 3% in comparison of r.
|
کلیدواژه
|
Energy dependence; H_p(10) uncertainty; ICRP trumpet graph; TLD-100H
|
آدرس
|
Universiti Kebangsaan Malaysia, Faculty of Science and Technology, School of Applied Physics, Malaysia, Universiti Kebangsaan Malaysia, Faculty of Science and Technology, School of Applied Physics, Malaysia, Universiti Kebangsaan Malaysia, Faculty of Science and Technology, School of Applied Physics, Malaysia
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|