SCREEN-PRINTED POTASSIUM ION SENSOR FABRICATED FROM PHOTOCURABLE AND SELF-PLASTICIZED ACRYLIC FILM
|
|
|
|
|
نویسنده
|
Alva Sagir ,Heng Lee Yook ,Ahmad Musa
|
منبع
|
journal of physical science - 2006 - دوره : 17 - شماره : 2 - صفحه:141 -150
|
چکیده
|
Potassium ion sensor based on photocured poly(n-butyl acrylate) membrane has been fabricated on screen printed electrode (spe) through a photocuring technique. the advantage of the fabrication method allows miniatured but two-dimensional sensor that is disposable to be designed. the potassium ion sensor yielded a nernstian response, i.e., 58.5 ± 0.5 mv/decade when the potassium ion concentrations were 10–1–10–5m. the limit of detection was close to (2.5 ± 0.57) × 10–6 m. the potassium ion spe also demonstrated little interference from cations such as ca2+, mg2+, li+, and na+ and good reproducibility with relative standard error of 0.81% (n = 4). the sensors were applied to the analysis of potassium in urine and the results were compared with that of a conventional technique such as flame photometry
|
کلیدواژه
|
Potassium ion sensor ,plasticizer free acrylic membtane ,urine analysis ,Ag/AgCI screen-printed electrode ,photopolymerization
|
آدرس
|
Universiti Kebangsaan Malaysia, Faculty of Science and Technology, School of Chemical Sciences and Food Technology, Malaysia, Universiti Kebangsaan Malaysia, Faculty of Science and Technology, School of Chemical Sciences and Food Technology, Malaysia, Universiti Kebangsaan Malaysia, Faculty of Science and Technology, School of Chemical Sciences and Food Technology, Malaysia
|
پست الکترونیکی
|
yhl1000@pkrisc.cc.ukm.my
|
|
|
|
|
|
|
|
|
|
|
|