Fa   |   Ar   |   En
   Highly Selective Perchlorate Membrane Electrode Based on Synthesized Schiff Base Complex  
   
نویسنده Mazloum-Ardakani Mohammad ,Mazidi Roya ,Mashhadizadeh Mohammad Hossein ,Naeimi Hossein ,Rabiei Kobra
منبع iranian journal of chemistry and chemical engineering - 2010 - دوره : 29 - شماره : 3 - صفحه:123 -132
چکیده    In this work we introduce a practical use of synthesized schiff base metal complex in sensing application. a new perchlorate-selective electrode based on the complex n, n'-4- trophenylmethanebis-( salicylaldiminato) copper (ii) [cul] as the membrane carrier was developed. the electrode exhibits a good nernstian slope of -55.0 ± 1.0 mv/decade and a linear range of 7.0 ×10 ^-7 - 1.0 ×10^-1m for perchlorate. the limit of detection is 3.0 ×10^ -7 m . it has a fast response time <15s and can be used for more than three months. the selective coefficients of perchlorate-selective electrode were determined by the fixed interference method (fim) and fixed primary ion method (fpm), and could be used in the ph range 4.0-9.0. it was employed as an indicator electrode for direct determination of perchlorate in real samples.
کلیدواژه Schiff base ,Perchlorate-selective electrode ,Potentiometry
آدرس yazd university, Faculty of Science, Department of Chemistry, ایران, payame noor university, Department of Chemistry, ایران, kharazmi university (university of tarbiat moallem), Department of Chemistry, ایران, university of kashan, Faculty of Chemistry, ایران, university of kashan, Faculty of Chemistry, ایران
پست الکترونیکی mazloum@yazduni.ac.ir
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved