A shift to administration of optimal dose of 18f-fdg between 4 and 5 mbq/kg from the current practice of higher doses potentially yields a reasonable-to-excellent pet image. for this purpose, whole-body mip images of 32 patients (23 men, 9 women, age 51.9 ± 13.7 years), administered with 18f-fdg (activity 5.3 ± 0.5 mbq/kg, 45 minutes uptake time) for whole-body pet/ct examinations, were evaluated. image quality was assessed visually by two radiologists using a three-point scoring scale: poor, reasonable and excellent. the interobserver agreement revealed a kappa value higher than 0.7. therefore, the utilisation of 18f-fdg dose between 4 and 5mbq/kg is considered an optimum dose for whole-body pet/ct examination.
|
|
کلیدواژه
|
18F-FDG ,PET/CT ,PET image quality ,optimum dose ,administered dose/body weight (kg) ,MIP
|
|
آدرس
|
Universiti Putra Malaysia, Centre for Diagnostic Nuclear Imaging, Malaysia, Universiti Putra Malaysia, Centre for Diagnostic Nuclear Imaging, Malaysia, Universiti Putra Malaysia, Centre for Diagnostic Nuclear Imaging, Malaysia, Universiti Putra Malaysia, Centre for Diagnostic Nuclear Imaging, Malaysia
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|