Fa   |   Ar   |   En
   Determination of Primary Electron Beam Parameters in A Siemens Primus Linac Using Monte Carlo Simulation  
   
نویسنده Seifi Makrani Danial ,Hasanzadeh Hadi ,Allahverdi Pourfallah Tayyeb ,Ghasemi Arash ,Jadidi Majid ,Babapour Hamed
منبع Archives Of Advances In Biosciences - 2015 - دوره : 6 - شماره : 1 - صفحه:75 -79
چکیده    The monte carlo method can be used to describe any technique that approximates solutions to quantitative problems through statistical sampling. this method is considered to be the most accurate method for dose calculation in radiotherapy. for complete modeling of a linear accelerator, it is required that the manufactured information covers all data, but some data such as primary electron energy must be indicated. the purpose of this study was to determine the best primary electron energy for 15 mv photon beam with varying the energy and fwhm. a monte carlo model for photon-beam output of a siemens primus linear accelerator was validated by plotting the energy spectrum of photon beam and calculating the percentage depth dose (pdd) and beam profiles for 10×10 cm2 field. square 10×10 cm² field was validated by measurements in water by a farmer chamber. linac head simulation was performed with beamnrc and dose calculation and 3d dose file were produced by dosxyznrc. the results were analyzed using matlab. it was found a good agreement between calculated pdd and beam profile for 15 mv photon beam using monte carlo simulation with primary electron energy of 11 mv and fwhm of 0.4 with maximum dose difference of 1.2% in pdd curves. in conclusion, using primary electron energy of 11 mv and fwhm of 0.4 has very good accuracy in calculating of dose distribution for 15 mv photon beam and it can be considered as a promising method for patient dose calculations.
کلیدواژه Beamnrc; Monte Carlo; Linac Simulation; Beam Parameters; Fwhm
آدرس Semnan University Of Medical Sciences, Faculty Of Medical Sciences, Department Of Medical Physics, ایران, Semnan University Of Medical Sciences, Cancer Research Center, Department Of Medical Physics, ایران, Mazandaran University Of Medical Sciences, Faculty Of Medicine, Department Of Biochemistry And Biophysics, ایران, Mazandaran University Of Medical Sciences, Faculty Of Medicine, Department Of Radiology And Radiation Oncology, ایران, Semnan University Of Medical Sciences, Faculty Of Medical Sciences, Department Of Medical Physics, ایران, Guilan Oncology Center (Goc), ایران
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved