Validation of Datascope Accutorr PlusTM using British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) protocol guidelines
|
|
|
|
|
نویسنده
|
Khawaja Rajab Ali ,Qureshi Riaz ,Mansure Abdul Hay ,Yahya Mohammad Emran
|
منبع
|
journal of the saudi heart association - 2010 - دوره : 22 - شماره : 1 - صفحه:1 -5
|
چکیده
|
To assure the accuracy and reliability of blood pressure measurement by noninvasive blood pressure monitor using datascope accutorr plus (paramus, nj, usa) against mercury manometer, among adult male participants. method: eighty participants from a family physician’s office at a teaching hospital were recruited. one hundred and sixty measurements of blood pressure were performed according to bhs technique protocol. statistical analysis: descriptive analysis was done according to the aami and bhs protocol guidelines. the limits of agreement between the device and the standard were plotted using the method of bland and altman plot. results: the mean difference ± sd between the datascope acutorr plus and observer was 2.7 ± 5.2 mm hg and 1.5± 3.26 mm hg for systolic and diastolic blood pressure, respectively. datascope acutorr plus obtained a/a grading for both systolic and diastolic blood pressure.
|
کلیدواژه
|
AAMI; BHS; BP devices; Datascope Acutorr plus; Validation
|
آدرس
|
King Saud University, King Khalid University Hospital, Department of Family and Community Medicine, Saudi Arabia, King Saud University, King Khalid University Hospital, Department of Family and Community Medicine, Saudi Arabia, King Saud University, King Khalid University Hospital, Department of Family and Community Medicine, Saudi Arabia, King Saud University, King Khalid University Hospital, Department of Family and Community Medicine, Saudi Arabia
|
پست الکترونیکی
|
rajab99@hotmail.com, rajabali999@yahoo.com
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|