Fa   |   Ar   |   En
   Computation of the full energy peak efficiency of an hpge detector using a new compact simulation analytical approach for spherical sources  
   
نویسنده el-khatib a.m. ,gouda m.m. ,badawi m.s. ,nafee s.s. ,el-mallah e.a.
منبع journal of engineering science and technology - 2013 - دوره : 8 - شماره : 5 - صفحه:623 -638
چکیده    The full energy peak efficiency of hpge detector is computed using a new analytical approach. the approach explains the effect of self-attenuation of the source matrix,the attenuation by the source container and the detector housing materials on the detector efficiency. the experimental calibration process was done using radioactive spherical sources containing aqueous 152eu radionuclide which produces photons with a wide range of energies from 121 up to 1408 kev. the comparison shows a good agreement between the measured and calculated efficiencies for the detector using spherical sources. © school of engineering,taylor's university.
کلیدواژه Full-energy peak efficiency; HPGe detectors; Self-attenuation; Spherical sources
آدرس physics department,alexandria university, Egypt, physics department,alexandria university, Egypt, physics department,alexandria university, Egypt, physics department,alexandria university, Egypt, physics department,alexandria university, Egypt
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center