   Development of a Doppler broadening positron spectroscopy setup and relative positron-electron momentum distribution measurement for commercial Al  
   
نویسنده Mehmandoost-Khajeh-Dad A. A. ,Mahjour-Shafiei M. ,Khaghani M.
منبع journal of theoretical and applied physics - 2011 - دوره : 4 - شماره : 4 - صفحه:9 -12
چکیده    Development of a doppler broadening positron spectroscopy setup is reported. the setup consists of an hpge detector with energy resolution of 2.1 kev at 411.8 kev in coincidence with a nai detector. the stability of the electronic devices was extensively studied. finally, the relative positron-electron momentum distribution for commercial al was measured. the comparison between our data and other’s data reveals a very good agreement.
کلیدواژه Positron annihilation; Doppler broadening; coincidence setup; system stability
آدرس university of sistan and baluchestan, Department of Physics, ایران, university of tehran, Department of Physics, ایران, university of sistan and baluchestan, Department of Physics, ایران