High frequency electromagnetic interference reduction on the squid attached in pulse tube cryocooler
|
|
|
|
|
نویسنده
|
Eshraghi M. J. ,Cho M. ,Kim J. Mok ,Sasada I.
|
منبع
|
journal of theoretical and applied physics - 2009 - دوره : 3 - شماره : 2 - صفحه:33 -36
|
چکیده
|
Measurements of the high frequency electromagnetic interference (emi) of a pulse tube cryocooler are made in atypical laboratory environment using laboratory-made current transformers. the inverter makes high frequencyemi which could cause severe interference with squid operation inside the pulse tube cryocooler. by coveringthe power lines, which include a neutral line, of the valve motor with a copper mesh cover, as a ground line, oneorder decrease in the magnitude of the emi noise is successfully obtained. this paper describes one example toshow how to observe and reduce conducted high frequency noise currents in a precision experimental system.
|
کلیدواژه
|
Electromagnetic interference ,PT cryocooler ,SQUIDs ,Noise measurement
|
آدرس
|
Kyushu University, Department of Applied Science for Electronics and Materials, Japan, Kyushu University, Department of Applied Science for Electronics and Materials, Japan, Korea Research Institute of Standards and Science, Korea, Kyushu University, Department of Applied Science for Electronics and Materials, Japan
|