patient-specific dosimetric validation of pelvic imrt and vmat plans using a custom-designed female pelvic phantom and arc-check qa system
|
|
|
|
|
|
|
|
نویسنده
|
katake ajay ,kumar lalit ,usadadiya hetvi ,haraniya kruti ,basandrai deepak
|
|
منبع
|
iranian journal of medical physics - 2025 - دوره : 22 - شماره : 6 - صفحه:395 -404
|
|
چکیده
|
Introduction: to ensure the accuracy of advanced radiotherapy techniques in the therapy of female pelvic malignancies using radiotherapy, comprehensive quality assurance protocols and patient-specific dosimetry was performed using a custom-designed female pelvic (cdfp) phantom and arc-check.material and methods: volumetric modulated arc therapy (vmat) and intensity modulated radiation therapy (imrt) plans originally developed for patients with cervical cancer were transferred to the cdfp phantom and arc-check. the doses for patient-specific quality assurance (psqa) of advanced techniques were calculated using different algorithms in treatment planning system (tps). to acquire dose measurements, a 0.600 cc ion chamber, sun nuclear 125c (snc125c) and a pinpoint 0.015 cc chamber were employed. the percentage deviation between the computed and administered dose was calculated. results: percentage discrepancies between planned and administered doses were analyzed. the maximum percentage deviation, mean ± standard deviation(sd) differences for 0.6 cc ion chamber and a pin-point 0.015 cc chamber were found to be -3.6%, -0.89% ± 1.59 and 3.51%, 1.10% ± 1.31 for the imrt treatment technique and -3.32%, -0.84% ± 1.53 and 3.6%, 1.25% ± 1.34 for the vmat treatment technique using aaa algorithms using 6mv beam for cdfp phantom respectively. the maximum of percentage deviation, mean ± sd differences for 0.6 cc ion chamber and a snc125c chamber were found to be -3.40%, -0.273% ± 2.161 and 3.107%, -0.028% ± 2.046 for cdfp phantom and -2.72%, -0.35% ± 1.41 and 2.74%, 0.197% ± 1.50 for arc-check using vmat treatment technique using 6mv beam and monte carlo algorithm respectively.conclusion: the inherent properties of anthropomorphic phantoms like the cdfp, such an anatomical fidelity and tissue equivalence, make them extremely useful in the quality assurance radiotherapy-programs. their application increases the accuracy of dosimetric validation and therefore, these phantoms are applicable in the use of high-accuracy treatment modalities such as imrt and vmat
|
|
کلیدواژه
|
cdfp arc ,check algorithms phantom quality assurance
|
|
آدرس
|
lovely professional university, school of chemical engineering and physical sciences, department of physics, india. n m virani wockhardt hospital, department of radiation oncology, india, max superspeciality hospital, department of radiation oncology, india, n m virani wockhardt hospital, department of radiation oncology, india, sir sayajirao general hospital, department of radiation oncology, india, lovely professional university, school of chemical engineering and physical sciences, department of physics, india
|
|
پست الکترونیکی
|
basandraideepak@yahoo.com, deepakbasandrai@gmail.com
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|