   suitability of chambers in relative dose measurement of small fields for accelerated photon delivered by a medical linear accelerator  
   
نویسنده purohit santunu ,haque meaze akm moinul ,kabir sm enamul ,rahman md. shakilur ,patwary kawchar ahmed
منبع iranian journal of medical physics - 2021 - دوره : 18 - شماره : 4 - صفحه:293 -299
چکیده    Introduction: using the small field in modern radiotherapy, the present study aimed at measuring the relative dosimetry (scattering factor, percentage depth dose (pdd), and profile of penumbra) with ionization (fc65-g, cc13, cc01) and diode (razor) chambers. material and methods: applying trs-398 in varian clinac™ ix-5982 for 6 mv photon beams, the conditions (pressure, temperature, direction, polarity) were kept the same for a set of field sizes (1 × 1, 2 × 2, 3 × 3, 4 × 4, 5 × 5, 7 × 7, and 10 × 10 cm^2), and relative dosimetry was performed at the north-east cancer hospital, sylhet, bangladesh. results: during the output factor measurement in small fields, the razor showed better results than cc13. taking cc01 as a standard in small fields, the data obtained from the study showed a good agreement with those of the previously published works. conclusion: razor, with extremely small active volume, was very much suited for small field dosimetry, except for pdds.
کلیدواژه clinical protocols ,scattering factors ,radiation dosimetry ,linac
آدرس university of chittagong, department of physics, bangladesh, university of chittagong, department of physics, bangladesh, north east cancer hospital, department of radiology, bangladesh, bangladesh atomic energy commission, secondary standard dosimetry laboratories (ssdls), bangladesh, comilla university, department of physics, bangladesh
پست الکترونیکی kapcu10@gmail.com
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved