suitability of chambers in relative dose measurement of small fields for accelerated photon delivered by a medical linear accelerator
|
|
|
|
|
نویسنده
|
purohit santunu ,haque meaze akm moinul ,kabir sm enamul ,rahman md. shakilur ,patwary kawchar ahmed
|
منبع
|
iranian journal of medical physics - 2021 - دوره : 18 - شماره : 4 - صفحه:293 -299
|
چکیده
|
Introduction: using the small field in modern radiotherapy, the present study aimed at measuring the relative dosimetry (scattering factor, percentage depth dose (pdd), and profile of penumbra) with ionization (fc65-g, cc13, cc01) and diode (razor) chambers. material and methods: applying trs-398 in varian clinac™ ix-5982 for 6 mv photon beams, the conditions (pressure, temperature, direction, polarity) were kept the same for a set of field sizes (1 × 1, 2 × 2, 3 × 3, 4 × 4, 5 × 5, 7 × 7, and 10 × 10 cm^2), and relative dosimetry was performed at the north-east cancer hospital, sylhet, bangladesh. results: during the output factor measurement in small fields, the razor showed better results than cc13. taking cc01 as a standard in small fields, the data obtained from the study showed a good agreement with those of the previously published works. conclusion: razor, with extremely small active volume, was very much suited for small field dosimetry, except for pdds.
|
کلیدواژه
|
clinical protocols ,scattering factors ,radiation dosimetry ,linac
|
آدرس
|
university of chittagong, department of physics, bangladesh, university of chittagong, department of physics, bangladesh, north east cancer hospital, department of radiology, bangladesh, bangladesh atomic energy commission, secondary standard dosimetry laboratories (ssdls), bangladesh, comilla university, department of physics, bangladesh
|
پست الکترونیکی
|
kapcu10@gmail.com
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|