   design and construction of an affordable phantom for electron density measurement and linearity tests of computed tomography systems  
   
نویسنده sefidbakht sepideh ,vani vardhan chatterjee ,ravanfar haghighi rezvan ,jalli reza ,chatterjee sabysachi
منبع iranian journal of medical physics - 2020 - دوره : 17 - شماره : 1 - صفحه:38 -47
چکیده    Introduction: the performance of computed tomography is routinely checked using phantoms, which are known as important diagnostic imaging tools. depending upon the aim of the study, different phantoms are designed, while trying to satisfy certain levels of diversities in their application. material and methods: the present study describes the construction of an inexpensive phantom designed for simultaneous measurements of 12 different samples. the body of the phantom, test tube holders, and test tubes were made of materials of low attenuation coefficient. body of the phantom was filled with water. test tubes filled with solutions of known chemical compositions were mounted on the test tube holders. the whole phantom was scanned at 80, 100, 120, 140 kvp to evaluate the performance of the ct system. using hounsfield unit (hu) data from these liquid samples of known electron density, the phantom was calibrated for electron density measurements. results: the system's accuracy and reproducibility were verified by measuring the hu values for some known materials. according to the results obtained from the experimental data with liquid samples, the accuracy of the water and noise was within ±3.2 hu and 0.6%, respectively. moreover, the image uniformity error was less than ±2 hu, and ct system's linearity for calibration was estimated with 99.9% confidence. conclusion: the present system gives satisfactory results with known samples and can be used with confidence for characterizing unknown materials.
کلیدواژه computed tomography phantom ,quality control attenuation ,photoelectric effect
آدرس shiraz university of medical sciences, medical imaging research centre, iran, indian institute of science, department of instrumentation and applied physics, india, shiraz university of medical sciences, medical imaging research center, iran, shiraz university of medical sciences, medical imaging research centre, , , iran, iran, , india
پست الکترونیکی chatsab99@gmail.com
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved