   assessment of x-ray crosstalk in a computed tomography scanner with small detector elements using monte carlo method  
   
نویسنده kavousi zahra ,karimian alireza ,jabbari iraj
منبع iranian journal of medical physics - 2018 - دوره : 15 - شماره : 3 - صفحه:169 -175
چکیده    Introduction: crosstalk is a leakage of x-ray or light produced in a matrix of x-ray detectors or array of photodiodes in one element to other elements affecting on image contrast and spatial resolution. in this study, we assessed x-ray crosstalk in a computed tomography (ct) scanner with small detector elements to estimate the effect of various parameters such as x-ray tube voltage, detector element sizes, scintillator material, impurities in the scintillator material, and the material of detector separators on x-ray crosstalk. materials and methods: this study was performed using monte carlo simulation. in the first step, x-ray tube and its energy spectrum at the energies of 80, 100, 120, and 140 kev were simulated and validated by using spekcalc and t-test. then, other important parts of ct scanner, namely filters, detectors, and grids were simulated. x-ray crosstalk between ct detectors was calculated in air and in the presence of water phantom (as a simulator of human body) to compare the effect of scattered photons. finally, the influence of some important parameters on x-ray crosstalk was evaluated. results: in ct scanner with small elements, when using phantom, crosstalk increases by 16-50%. using the lowest possible energies of x-ray, decreases the crosstalk up to 43% of its initial amount. furthermore coating a 10 or 20 μm layer of tungsten or lead on the detector separators, decreases the x-ray crosstalk significantly. conclusion: choosing the proper high voltage, detectors’ material and its dimensions, scintillator impurities and septa material can decrease x-ray crosstalk.
کلیدواژه computed tomography ,detector ,monte carlo method ,phantom
آدرس university of isfahan, faculty of engineering, department of biomedical engineering, iran, university of isfahan, faculty of engineering, department of biomedical engineering, iran, university of isfahan, faculty of new science and technologies, department of nuclear engineering, iran
پست الکترونیکی i_jabbari@ast.ui.ac.ir
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved