Precise Water Isotope Analysis in the Natural Concentration Range Using Off-Axis Integrated Cavity Output Spectroscopy (OA-ICOS)
|
|
|
|
|
|
|
|
نویسنده
|
asgary amir ,jaleh babak
|
|
منبع
|
international journal of optics and photonics - 2024 - دوره : 18 - شماره : 1 - صفحه:43 -50
|
|
چکیده
|
A compact sensor utilizing off-axis integrated cavity output spectroscopy at a wavelength of 1.389 μm was developed to measure the δ¹⁸o and δ²h isotopic ratios in liquid water samples within the natural concentration range. allan variance analysis indicated an optimal averaging time of approximately 21 s. the stability and long-term reproducibility of the laser system were assessed using 1-second averaging over 75 min. the minimum detectable absorption (mda) was determined to be approximately 8.4 × 10⁻⁴ hz⁻¹/² with a 17 s averaging time. to assess accuracy and reproducibility, the setup was tested within the natural concentration range (δ²h: -400 to 0‰ and δ¹⁸o: -54 to 0‰) using five standard samples, achieving an average accuracy and reproducibility of better than 2.5‰ for δ²h and 0.5‰ for δ¹⁸o.
|
|
کلیدواژه
|
Water Isotopes ,Natural concentration ,Laser Spectroscopy ,Allan deviation ,Sensitivity ,Accuracy ,Averaging
|
|
آدرس
|
bu-ali sina university, faculty of science, deprtment of physics, Iran. arak university, institute of molecular sciences and engineering, Iran, bu-ali sina university, faculty of science, deprtment of physics, Iran
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|